CHENG Yong-jun, LI De-tian. CALIBRATION OF EXTREMELY HIGH VACUUM IONIZATION GAUGE[J]. VACUUM AND CRYOGENICS, 2009, 15(4): 199-206.
Citation: CHENG Yong-jun, LI De-tian. CALIBRATION OF EXTREMELY HIGH VACUUM IONIZATION GAUGE[J]. VACUUM AND CRYOGENICS, 2009, 15(4): 199-206.

CALIBRATION OF EXTREMELY HIGH VACUUM IONIZATION GAUGE

  • Two inverted magnetron cold cathode ionization gauges and one hot cathode ionization gauge were calibrated against a(extremely high vacuum) XHV calibration apparatus over the range 10-9 Pa to 10-1Pa with gas of N2.The results showed that all gauges had better repeatability.Nevertheless,as for a whole measurement course,the correction factor and sensitivity were strongly pressure dependent.Therefore,to ensure accurate pressure measurement,it is necessary to calibrate these XHV ionization gauges accurately for reliability of the application in XHV.